WGZ-4000/2000/1000/500/200/100 Ratio Turbidimeter
Characteristics
Measurement of light scatter and decay of the insoluble particles suspended in water or transparent liquids;
Quantitative analysis of particle contents.
Specifications
WGZ-4000 | WGZ-2000 | WGZ-1000 | |||
Light source: | Tungsten halogen lamp 6V, 10W | ||||
Receiving element: | Silicon photocell | ||||
Measurement range NTU: | 0.00-50.0;50.1-200; (automatic range switchover) | 0.00-50.0;50.1-200; 200.1-2000 (automatic range switchover) | 0.00-50.0;50.1-200 200.1-1000 (automatic range switchover) | ||
Readings display: | LCD Touch Screen | ||||
Minimum reading | 0.001 | ||||
Allowable error of indication: | ≤±6% | ||||
Zero Drift: | ≤±0.3%FS | ||||
Stability of indication: | ≤±1%FS | ||||
Repeatability: | ≤0.5% | ||||
Sample bottle: | Φ 25 mm×95 mm | ||||
Data storage: | 1000 sets | ||||
Calibration modes: | 7 | 6 | 5 | ||
Sample volume: | 30--40 ml | ||||
Interface: | USB/U disk | ||||
Weight (Gross): | 9kg | ||||
Overall dimensions: | 430mm×430mm×300mm | ||||
WGZ-500 | WGZ-200 | WGZ-100 | |||
Light source: | Tungsten halogen lamp 6V, 10W | ||||
Receiving element: | Silicon photocell | ||||
Measurement range NTU: | 0.00-50.0;50.1-200; (automatic range switchover) | 0.00-50.0; 50.1-200; (automatic range switchover) | 0.00-50.0; 50.1-100 (automatic range switchover) | ||
Readings display: | LCD Touch Screen | ||||
Minimum reading | 0.001 | ||||
Allowable error of indication: | ≤±6% | ||||
Zero Drift: | ≤±0.3%FS | ||||
Stability of indication: | ≤±1%FS | ||||
Repeatability: | ≤0.5% | ||||
Sample bottle: | Φ 25 mm×95 mm | ||||
Data storage: | 1000 sets | ||||
Calibration modes: | 4 | 3 | 3 | ||
Sample volume: | 30--40 ml | ||||
Interface: | USB/U disk | ||||
Weight (Gross): | 9kg | ||||
Overall dimensions: | 430mm×430mm×300mm |